DS-25 DENSITOMETER
Calibrating DS-25 USB DENSITOMETER
Prepare LO-CAL (white) field of Calibration Tablet.
Make measurement of LO-CAL field of Calibration Tablet.
Do not lift Densitometer after measurement !
Wait about 8 - 10 second until hear continuous repeated sound.
.
After hear continuous repeated sound, Densitometer must be lift..
Prepare HI-CAL (black) field of Calibration Tablet.
Make measurement of HI-CAL field of Calibration Tablet.
After measurement is made, Densitometer must be lift..
End of Calibrating procedure. Densitometer is now calibrated.