Selected_Methods_Chapter37.pdf

(417 KB) Pobierz
196658805 UNPDF
CHAPTER 37
Selected Methods of Analysis
Chemistry is primarily an experimental science. This chapter presents a variety of
laboratory experiments, from classical titrations and gravimetry to instrumental
methods such as chromatography and spectroscopy. Detailed directions are given
for each experiment.
analyses. The methods have been chosen to introduce you to analytical tech-
niques that are widely used by chemists. For most of these analyses, the composi-
tion of the samples is known to the instructor. Thus, you will be able to judge how
well you are mastering these techniques.
Your chances of success in the laboratory will greatly improve if you take time
before you enter the laboratory to read carefully and understand each step in the
method and to develop a plan for how and when you will perform each step.
The discussion in this section is aimed at helping you develop efficient work
habits in the laboratory and also at providing you with some general information
about an analytical chemistry laboratory. Before you start an analysis, you should
understand the significance of each step in the procedure to avoid the pitfalls and
potential sources of error that are inherent in all analytical methods. Information
about these steps can usually be found in (1) preliminary discussion sections, (2)
earlier chapters that are referred to in the discussion section, and (3) the “Notes”
that follow many of the procedures. If, after reading these materials, you still do not
understand the reason for doing one or more of the steps in the method, consult
your instructor before you begin laboratory work.
The Accuracy of Measurements
In looking over an analytical procedure, you should decide which measurements
must be made with maximum precision, and thus with maximum care, as opposed
to those that can be carried out rapidly with little concern for precision. Generally,
measurements that appear in the equation used to compute the results must be per-
formed with maximum precision. The remaining measurements can and should be
made less carefully to conserve time. The words about and approximately are fre-
quently used to indicate that a measurement does not have to be done carefully. For
example, you should not waste time and effort to measure a volume to
0.02 mL
when an uncertainty of
0.5 mL or even
5 mL will have no discernible effect on
the results.
In some procedures, a statement such as “weigh three 0.5-g samples to the near-
est 0.1 mg” is encountered. Here, samples of perhaps 0.4 to 0.6 g are acceptable,
but their masses must be known to the nearest 0.1 mg. The number of significant
T his chapter contains detailed directions for performing a variety of chemical
196658805.007.png
37A An Introductory Experiment
1053
figures in the specification of a volume or a mass is also a guide to the care that
should be taken in making a measurement. For example, the statement “add 10.00
mL of a solution to the beaker” indicates that you should measure the volume care-
fully with a buret or a pipet, with the aim of limiting the uncertainty to perhaps
Time Utilization
You should study carefully the time requirements of the several unit operations
involved in an analysis before work is started. This study will reveal operations that
require considerable elapsed, or clock, time but little or no operator time. Examples
of such operations include drying a sample in an oven, cooling a sample in a desicca-
tor, or evaporating liquid on a hot plate. Efficient workers use such periods to perform
other operations or perhaps to begin a new analysis. Some people find it worthwhile
to prepare a written time schedule for each laboratory period to avoid dead time.
Time planning is also needed to identify places where an analysis can be inter-
rupted for overnight or longer, as well as those operations that must be completed
without a break.
Reagents
Directions for the preparation of reagents accompany many of the procedures.
Before preparing such reagents, be sure to check to see if they are already prepared
and available on a side shelf for general use.
If a reagent is known to be hazardous, you should plan in advance of the labo-
ratory period the steps that you should take to minimize injury or damage. Further-
more, you must acquaint yourself with the rules that apply in your laboratory for
the disposal of waste liquids and solids. These rules vary from one part of the coun-
try to another and even among laboratories in the same locale.
Water
Some laboratories use deionizers to purify water; others employ stills for this pur-
pose. The terms “distilled water” and “deionized water” are used interchangeably
in the directions that follow. Either type is satisfactory for the procedures in this
chapter.
You should use tap water only for preliminary cleaning of glassware. The
cleaned glassware is then rinsed with at least three small portions of distilled or
deionized water.
37A AN INTRODUCTORY EXPERIMENT
The purpose of this experiment is to introduce several of the tools, techniques, and
skills necessary for work in the analytical chemistry laboratory. The techniques are
considered one at a time, as unit operations. It is important to learn proper tech-
niques and to acquire individual skills before attempting additional laboratory
experiments.
37A-1 Using the Analytical Balance
Discussion
In this experiment, you will obtain the mass of five new pennies—first by deter-
mining the mass of each penny individually. Then you will determine the mass of
0.02 mL. In contrast, if the directions read “add 10 mL,” the measurement can be
made with a graduated cylinder.
196658805.008.png
1054
CHAPTER 37
Selected Methods of Analysis
all five pennies at once, remove one penny at a time, and calculate the individual
masses of the pennies by finding the difference. The pair of masses determined for
a particular penny by the two different methods should agree to within a few tenths
of a milligram. From the data, you will determine the mean and median values, the
standard deviation, and the relative standard deviation of the masses of the pennies.
You will then weigh an unknown aluminum cylinder and report the mass of this
unknown.
PROCEDURE
1. After you have been instructed in the use of the balance and have become famil-
iar with its use, obtain a set of pennies, an unknown aluminum cylinder, and a
pair of tweezers from the instructor.
2. Do not handle the pennies or the cylinder with your fingers; always use the
tweezers. If you are using a mechanical balance, be sure to have the balance in
the “off” or “complete arrest” position whenever removing anything from or
adding anything to the balance pan.
3. Before you begin to determine masses, zero your analytical balance carefully.
Select five pennies at random from the vial containing the pennies, and weigh
each penny on your balance. Enter the data in your laboratory notebook. Keep
track of the identity of each penny by placing each one on a labeled piece of
paper.
4. Check the zero setting on your balance. Place these same five pennies on the
balance pan, determine their total mass, and record it.
5. Remove one of the pennies from the balance, obtain the mass of the remaining
four, and record the mass.
6. Repeat this process, removing one penny at a time. Obtain the individual
masses by subtraction. This process is known as weighing by difference, which
is the way many mass determinations are done in the analytical laboratory.
7. Finally, check the zero on your balance, and find the mass of the unknown alu-
minum cylinder.
37A-2 Making Quantitative Transfers
Discussion
The following experiment is designed to provide experience in the correct use of
the volumetric flask.
PROCEDURE
1. Weigh a 50-mL beaker on a triple-beam balance or an appropriate electronic
top-loading balance.
2. Adjust the balance for an additional 0.4 g and add solid KMnO 4 to the beaker
until the beam is again balanced. If you have an electronic balance with a tare
function, depress the tare button to set the balance to zero. Then add KMnO 4
until the balance reads about 0.4 g. Note that chemicals should never be
returned to a stock bottle, as this may contaminate the bottle.
3. Dissolve the potassium permanganate in the beaker using about 20 mL of dis-
tilled water. Stir gently to avoid loss. This is nearly a saturated solution, and
some care is required to dissolve the crystals completely.
196658805.009.png 196658805.010.png 196658805.001.png
37A An Introductory Experiment
1055
4. Quantitatively transfer the solution to a 100-mL volumetric flask fitted with a
small funnel. To prevent solution from running down the outside of the beaker,
pour it down the stirring rod, and then touch the rod to the spout of the beaker to
remove the last drop. Add more water to the beaker, stir, and repeat the
procedure.
5. Repeat the procedure until no trace of the color of the permanganate remains in
the beaker. Note the number of washings that is required to quantitatively trans-
fer the permanganate from the beaker to the flask.
6. Rinse the last portion of solution from the stirring rod into the volumetric flask
with a stream of water from the wash bottle. Rinse the funnel and remove it.
Dilute the solution in the flask until the bottom of the meniscus is even with the
graduation mark. Stopper, invert, and shake the flask. Return it to the upright
position, and allow the air bubble to return all the way to the top of the neck.
7. Repeat until the solution is completely homogeneous; about 10 inversions and
shakings are required. Save the solution for Part 37A-3.
37A-3 Delivering an Aliquot
Discussion
Whenever a buret or pipet is used to deliver a measured volume of solution, the liq-
uid it contains before measurement should have the same composition as the solu-
tion to be dispensed. The following operations are designed to illustrate how to
rinse and fill a pipet and how to deliver an aliquot of solution.
PROCEDURE
1. Fill a pipet with the solution of potassium permanganate and let it drain.
2. Draw a few milliliters of distilled water from a 50-mL beaker into the pipet,
rinse all internal surfaces of the pipet, and discard the rinse solution. Do not fill
the pipet completely; this is wasteful, time-consuming, and inefficient. Just
draw in a small amount, tilt the pipet horizontally, and turn it to rinse the sides.
3. Determine the minimum number of such rinsings required to completely
remove the permanganate color from the pipet. If your technique is efficient,
three rinsings should be enough.
4. Again fill the pipet with permanganate solution, and proceed as before. This
time determine the minimum volume of rinse water required to remove the
color by collecting the rinsings in a graduated cylinder. Less than 5 mL are
enough with efficient technique. In the rinsing operations, was the water in the
50-mL beaker contaminated with permanganate? If a pink color shows that it
was, repeat the exercise with more care.
5. As a test of your technique, ask the laboratory instructor to observe and com-
ment on the following operation: Rinse a 10-mL pipet several times with the
solution of potassium permanganate you prepared.
6. Pipet 10 mL of the permanganate solution into a 250-mL volumetric flask.
7. Carefully dilute the solution to volume, trying to mix the contents as little as
possible.
8. Mix the solution by repeatedly inverting and shaking the flask. Note the effort
that is required to disperse the permanganate color uniformly throughout the
solution.
9. Rinse the pipet with the solution in the volumetric flask. Pipet a 10-mL aliquot
of the solution into a conical flask.
196658805.002.png 196658805.003.png
1056
CHAPTER 37
Selected Methods of Analysis
37A-4 Calibrating a Pipet
Discussion
The proper manual technique for calibrating an analytical transfer pipet is readily
learned with practice, care, and attention to detail. With the possible exception of
mass determinations, this experiment has the potential of being the most accurate
and precise set of measurements that you will ever make.
PROCEDURE
1. Clean a 10-mL pipet. When a pipet, buret, or other piece of volumetric glass-
ware is cleaned properly, no droplets of reagent remain on the internal surfaces
when they are drained. This is very important for accurate and reproducible
results. If reagent adheres to the inside of a pipet, you cannot deliver the nomi-
nal volume of the pipet. If you clean a pipet or any other glassware with alco-
holic KOH, use the bottle of cleaning solution only inside the sink and rinse it
off thoroughly before returning it to the shelf. Do not put the bottle of cleaning
solution directly on a bench top; it may ruin the surface. The solution is very
corrosive. If your fingers feel slippery after use, or if some part of your body
develops an itch, wash the area thoroughly with water.
2. Obtain a pipetting bulb, a 50-mL Erlenmeyer flask with a dry stopper, a 400-
mL beaker of distilled water equilibrated to room temperature, and a
thermometer.
3. Determine the mass of the flask and stopper and record it to the nearest 0.1 mg.
Do not touch the flask with your fingers after this weighing. Use tongs or a
folded strip of waxed paper to manipulate the flask.
4. Measure and record the temperature of the water.
5. Pipet 10.00 mL of the distilled water into the flask using the technique
described on page 45. Stopper the flask, determine the mass of the flask and the
water that it contains, and record the mass.
6. In the same way, add a second pipet of water to the flask; remove the stopper
just before the addition. Replace the stopper, and once again determine and
record the mass of the flask and the water. Following each trial, determine the
mass of water added to the flask by the pipet.
7. Repeat this process until you have determined four consecutive masses of water
that agree within a range of 0.02 g. If the determinations of the mass of water
delivered by the pipet do not agree within this range, your pipetting technique
may be suspect. Consult your instructor for assistance in finding the source of
the error, and then repeat the experiment until you are able to deliver four con-
secutive volumes of water with the precision cited.
8. Correct the mass for buoyancy as described on page 27, and calculate the vol-
ume of the pipet in milliliters.
9. Report the mean, the standard deviation, and the relative standard deviation of
the volume of your pipet. Calculate and report the 95% confidence interval for
the volume of your pipet.
37A-5 Reading Buret Sections
Discussion
The following exercise will give you practice in reading a buret and confirming the
accuracy of your readings.
196658805.004.png 196658805.005.png 196658805.006.png
Zgłoś jeśli naruszono regulamin