appxav1.pdf

(402 KB) Pobierz
Cristalli, C., Manzoni, A. “Basics of Blood Gas Instrumentation.”
The Biomedical Engineering Handbook: Second Edition.
Ed. Joseph D. Bronzino
Boca Raton: CRC Press LLC, 2000
A
A PPENDIX
Basics of Blood Gas
Instrumentation
A.1
pH and PCO
Electrochemical Sensors
2
A.2
PO
Electrochemical Sensors of Dissolved Oxygen
2
A.3
Equivalent Electronic Circuit for a Potentiometric
Electrochemical Cell
Cristina Cristalli
A.4
General Aspects of a Mearuing Device
A.5
Factors which Influence the Measurements
Instrumentation Laboratory
A.6
New Generations of pH and Blood Gas Sensors
Angelo Manzoni
A.7
Continuous Monitoring of Blood Gases
Instrumentation Laboratory
A.8
Optical Sensors in Blood Gas Analyzers
Modern anesthesiology and intensive care, neonatal care, and emergency medical facilities depend heavily
upon the frequent measurements of a selected group of critical care analytes in blood. Tests recognized
as being essential for patient management include pH and blood gas (O
partial pressures) and
electrolytes (sodium, potassium, and calcium) measurements. Clinical care clinicians use information
from blood gas reports in a variety of ways for patient interpretation and to establish plans for patient
therapy. The measurement of pH and PCO
and CO
2
2
, for example, represents any disturbances in acid–base
balance, and the calculation of the bicarbonate concentration from the Henderson–Hasselbalch equation
gives the most complete and consistent model of acid–base physiology for use in infants, children, and
adults with a variety of disorders. Several efforts, with different levels of success, have been made in order
to determine the primary variables to be measured directly from whole blood in critical and emergency
situations. In fact, in a different way from the clinical chemistry routine, some analytes must be deter-
mined quickly
2
at the patient’s bedside or emergency site without modifying the sample. Instru-
mentation for these measurements should be as effective as the
in vitro
in vivo
monitoring and should be capable
of making the same determinations as the clinical laboratory
.
Electrochemical sensors have been demonstrated to be the most appropriate way to achieve the
objective of the measurement of blood components without any sample treatment. In 1905, the first pH
glass electrochemical determination was demonstrated, and the first practical pH glass electrode for
anaerobic blood pH determination was the capillary electrode of McInnes and Belcher introduced in
early 1930. Blood pH was not considered interesting from a clinical point of view until 1952, when a
polio epidemic was seen in Copenhagen and patients required artificial ventilation. Astrup adopted the
measurement of pH and CO
to teach students how to change the
patient’s ventilation towards a more normal level. In August 1954, Stow described his concepts and
construction of a CO
content for calculation of the PCO
2
2
electrode capable of use in measuring blood PCO
, sensitive to CO
but insensitive
2
2
2
to acids and alkali in fluids of known PCO
. In 1953, Clark presented the first polyethylene-covered
oxygen electrode. This marked an historic turning point in blood gas analysis and all of respiratory
physiology. In 1959, the first example of a three-electrode blood gas apparatus containing a Beck-
2
© 2000 by CRC Press LLC
809259411.006.png 809259411.007.png
 
electrode, and a McInnes/Belcher pH electrode
was described. After a few years (early 1960s), Thomas Ross, the founder of Instrumentation Laboratory,
introduced the first commercial three-electrode system in the market (IL 113). Further development
brought an interesting level of automation to blood gas analysis, satisfying user demands at the beginning
of the 1970s when Radiometer introduced the first automated blood gas analyzer in 1973, ABL 1. Corning
Diagnostics (now Bayer), Nova Biomedical, AVL, and Eschweiler are companies that have used the blood
gas analyzer before the introduction of the portable blood gas analyzer by Mallinckrodt Sensor Systems
(now IL), iSTAT (now Abbot/HP) and Diametrics. The portable blood gas analyzer was developed in
order to bring the analyzer closer to the patients, the so-called Near Patient or Point of Care testing.
Sensor technology has notably improved in the development and manufacturing of sensors for the
direct determination of components of diagnostic interest such as sodium, potassium, calcium, lithium,
magnesium, and chloride for the electrolytes, and glucose, lactate, urea, and creatinine for the substrates.
This makes it possible to not only determine the blood pH and gases, but to do this together to other
analytes so that a complete diagnostic profile of the patient can be obtained.
In general, electrochemical sensors transform the information arising from the concentration of
chemical species in solution or the interaction or reaction of a substance with its substrate to an electrical
signal of an equivalent and measurable quantity, such as a potential, current, or charge transfer. Biomed-
ical sensors work as an interface between the biological system and the electrical one, without modifying
the characteristics of the two systems, guaranteeing the reliability of the resulting information. This
appendix will go over the basic concepts of blood gas instruments describing the principles for the
measurements of the signals coming from different sensors and the main criteria for data acquisition
and reduction. It will conclude with an overview of the new trends in sensor technologies.
man/Clark oxygen electrode, a Stow/Severinghaus CO
2
A.1
pH and PCO 2 Electrochemical Sensors
Electrochemical potentiometric systems constitute the main category of sensors for the determination
of ionic and molecular species in blood, such as pH, PCO
, and electrolytes. pH and sodium
potentiometric sensors are based on different glass membrane compositions obtained through the mix-
ture of metal–alkali oxides. Potassium, calcium, chloride, lithium, and magnesium potentiometric sensors
are based on polymeric matrix membranes containing the electroactive material that makes the mem-
brane selective for the determination of the respective ionic species in the blood. The potentiometric
sensors obtained by incorporating these membranes in their structures are called Ion Selective Electrodes
(ISE). They make possible the definition of an electrochemical system in which a potential difference is
seen at the sensor (ISE) when it is in contact with the sample to be measured (electrolytic solution). The
resulting electrical signal is dependent on the concentration, or more correctly on the ionic activity, of
the particular ionic species in the sample.
Figure A.1 s chematically illustrates the main components of a potentiometric pH sensor and its
reference electrode. In order to be able to measure the electrical signal present at the sensor/sample
interface, the Ion Selective Electrode (indicator electrode) should be connected to one of the two input
terminals of the potential measuring device, the Voltmeter, together with a second electrode, the reference
electrode, connected to the second input terminal of the Voltmeter. The structure and characteristic of
the reference electrode that is required to complete the potentiometric measuring system guarantees that
the change of the measured electrical signal depends only on the concentration of the ionic species in
solution interacting with the ISE membrane surface. The relationship between the ionic concentration
of the species in solution (a) and the measured electrical signal (E) is represented by the Nernst equation:
, PO
2
2
E = E° + K log a
where E° is the standard potential of the electrochemical system including the reference electrode potential
and K is the slope or the sensitivity of the electrochemical system expressed in mV/log a.
© 2000 by CRC Press LLC
809259411.008.png 809259411.001.png
 
FIGURE A.1
pH sensor and reference electrode connected to the measuring device.
FIGURE A.2
PCO
sensor structure.
2
In the actual blood gas analytical instrumentation, the determination of CO
partial pressure (PCO
)
2
2
is performed with a potentiometric sensor, the so-called Stowe-Severinghaus sensor.
The experimental setup represented in Fig. A.2 s hows the schematic diagram of the electrochemical
system for the measurement of a molecular species in solution such as CO
.
2
that enters the blood remains in plasma; the balance enters the red cell. A small
amount of the plasma CO
Only 5% of the CO
2
forms carbonic acid and the balance remains as dissolved CO
(dCO
) which
2
2
2
exerts a pressure (Henry’s law) measured as PCO
, in equilibrium in the sample, diffuses
through a gas permeable membrane, and changes the pH of the thin layer of bicarbonate-based electrolyte.
. The dCO
2
2
H +
(A.1)
dCO 2
CO 2
+
H 2 OH 2 CO 3
↔↔
+
HCO 3
The pH variation of the bicarbonate solution, measured by a pH sensor, is directly proportional to
the logarithm partial pressure of the CO
dissolved in the sample.
The logarithmic output of these sensors creates problems of inaccuracy because small uncertainties
in the measurement of the potential can cause significant variations in the determination of the concen-
tration of the species to measure. Precision and accuracy in the measurement can be reached with
2
© 2000 by CRC Press LLC
809259411.002.png
FIGURE A.3
Schematic view of the structure of an amperometric oxygen sensor.
appropriate calibration procedures. The calibration procedures should be designed in order to minimize
the influence of the reference electrode and the interference of other ionic species present in the sample
on the measured potential. pH sensors are calibrated with buffer solutions having pHs referred to as
NIST primary standards: phosphate buffers at pH = 7.392 and 6.839 at 37°C, with an ionic strength of
0.1 mole/Kg. Gas sensors (PCO
) are generally calibrated with humidified gas mixtures taking
into account the barometric pressure and temperature of each gas as well as the vapor pressure of water.
and PO
2
2
A.2
PO
Electrochemical Sensors of Dissolved Oxygen
2
) mea-
surement is made by an electrochemical amperometric sensor introduced by Clark in the beginning of
the 1950s. Most of the total oxygen content in blood is bound to hemoglobin. The remaining is dissolved
in plasma, and PO
In the majority of commercially available blood gas analyzers, the oxygen partial pressure (PO
2
(the measured quantity) is directly related to the concentration of O
dissolved in
2
2
plasma. Figure A.3 s hows a schematic diagram of a typical amperometric oxygen sensor.
The amperometric sensor for the determination of the partial pressure of molecular oxygen in an
aqueous solution depends on the chemical reduction of oxygen at a metallic or other conducting surface
(working electrode) when a potential difference between the reference electrode and the working electrode
is applied by an external voltage source. This constitutes the difference between amperometric and
potentiometric sensors; an external energy source drives an electrochemical reaction of the analyte
resulting in a current that is proportional to the analyte partial pressure in the amperometric case, while
potentiometric measurements such as pH and PCO
take place without any applied voltage and no
2
current flow in the measuring circuit.
The oxygen sensor measures the current originated during the reduction of oxygen according to the
reaction:
4e
4H +
(A.2)
O 2
+
+
H 2 O
to a platinum electrode to which a proper voltage to support the reaction (ca. –0.7 V) vs. an Ag/AgCl
reference/anode electrode) has been applied.
The platinum sensor is covered by an oxygen permeable membrane that protects it from a possible
poisoning caused by the presence of different substances present in the sample, such as the proteins, and
controls the oxygen diffusion at the platinum surface.
Silver/silver chloride (Ag/AgCl) is the most common reference electrode used in the amperometric
oxygen sensor. In fact, the chloride ions contained in the electrolyte reacts with the silver anode to form
© 2000 by CRC Press LLC
809259411.003.png 809259411.004.png 809259411.005.png
 
Zgłoś jeśli naruszono regulamin